ReLeaf RESTORATIVE Osnovni komplet za pripremu

ReLeaf-RESTORATIVE-Basic-Setup-Kit-product

Specifikacije

  • Materijal: Meki, fleksibilni termoplastični elastomer
  • 100% bez BPA i lateksa
  • Namjena: Restorativni stomatološki postupci

OPIS PROIZVODA/NAMJENA

ReLeaf® Restorative je uređaj za stomatološko usisavanje velikog volumena bez upotrebe ruku koji pomaže stomatologima pri evakuaciji, izolaciji, retrakciji i održavanju suhog polja tokom općih stomatoloških, restorativnih i higijenskih postupaka. ReLeaf® je namijenjen kliničkoj upotrebi od strane stomatologa.
Usnik Leaf izrađen je od mekog, fleksibilnog termoplastičnog elastomera koji je 100% bez BPA i lateksa. Nisu utvrđene nikakve zdravstvene opasnosti pri upotrebi ovog materijala.
Crijeva ReLeaf® izrađena su od silikonske gume (silikona i silicijum dioksida).

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (1)

INSTALACIJA I UPOTREBA OD STRANE PACIJENTA

  1. Fold the two Leaf halves together by pushing together the small protruding knob into the opposing hole at the Restorative Leaf base. Insert the larger knob into the opposing hole at the Leaf tip. Squeeze the two halves of the Leaf together to ensure each side securely attaches to the other. Attach the Leaf to the Restorative Adapter on the Short Hose (8” Quick Disconnect Hose) with the smaller leaf face facing away from the hose. Ensure the Leaf is securely attached.
  2. Insert the ReLeaf® Bite Block onto the peg closest to the Leaf on the Restorative Adapter. Making sure the concaved slope is in contact with the Restorative Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (2)
  3. Attach the free end of the
    Short Hose to the Quick Disconnect Adapter on the Long Hose (48” HVE Attachment Hose). Insert the coupler at the other end of the Long Hose securely into an open HVE port. It may be necessary to wet the
    O- ring inside the unused port so the Adapter can slide in easily.
  4. Place the ReLeaf® Bite Block on the opposing arch you are working on. While having the Leaf sweep in front of the tongue and wrap around the retromolar area to the opposing cheek side.
  5. When transitioning between patients, a cleaned and sterilized Bite Block, Restorative Adapter, and Short Hose must be utilized before attaching a new Leaf.ReLeaf-RESTORATIVE-Basic-Setup-Kit- (3)

KONTRAINDIKACIJE
Nema poznatih kontraindikacija.

MJERE PREDOSTROŽNOSTI

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

STERILIZACIJA, ODLAGANJE I SKLADIŠTENJE

All gray and white components of the ReLeaf™ system are compatible with standard dental sterilization protocols. These parts can be: wiped with dental-approved disinfectant wipes or sprays, run through dental atomizer, processed through an ultrasonic bath, run through a thermal washer, and run through autoclave cycle per sterilization cycle table.

  • Listovi: Za jednokratnu upotrebu, za jednokratnu upotrebu
    Jednokratni listovi namijenjeni su samo za jednokratnu upotrebu. Zbog prirode materijala i dizajna listova, ne smiju se sterilizirati i ponovo koristiti. Usnici od listova moraju se pravilno odložiti u spremnik za klinički otpad kako bi se spriječio prijenos zaraznih patogena.
  • Bite Block, Restorative Adapter and Short Hose
    100% Autoclavable and Reusable.
    Before placement in an ultrasonic cleaner, rinse the Bite Block, Restorative Adapter and Short Hose with water and brush both external and lumen surfaces with a soft bristle brush for the exterior surface and a soft bristle lumen brush or pipe cleaner for the interior surfaces. Place in an ultrasonic unit filled with enzymatic cleaner. (note: follow instructions with enzymatic cleaner for proper mixing, exposure time, etc., per manufacturer instructions). Rinse with water, place in an autoclavable pouch. Run product through a wrapped / pouched cycle in the autoclave.
  • Long Hose, Quick Disconnect Adapter and HVE Adapter
    100% Autoclavable and Reusable.
    The Long Hose should stay connected to the HVE until ready for routine evacuation cleaning with an enzymatic evacuation cleaner. Running evacuation cleaners through the hose to clean out any debris at least once per week is recommended. It is not necessary to run evacuation cleaners through the hose between each patient. If running through the autoclave, follow instructions listed above for Bite Block, Restorative Adapter & Short Hose.

Skladištenje
Molimo Vas da Releaf® čuvate na sobnoj temperaturi.
The Long Hose and Short Hose are designed to be held in the HVE Adapter when unused. (See image)

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (4)

VALIDIRANI CIKLUSI STERILIZACIJE

Autoklav sa gravitacionim pomeranjem

  • TEMPERATURA: 121°C/250°F
  • MINIMALNO VRIJEME IZLOŽENOSTI: 30 minuta
  • MINIMALNO VRIJEME SUŠENJA: 30 min
VALIDIRANI CIKLUSI STERILIZACIJE
Autoklav sa gravitacionim pomeranjem Autoklav prije vakuuma
TEMPERATURE: 121°C/250°F TEMPERATURE: 132°C/270°F
MINIMALNO VRIJEME IZLOŽENOSTI: 30 minuta MINIMALNO VRIJEME IZLOŽENOSTI: 4 minuta
MINIMALNO VRIJEME SUŠENJA: 30 min MINIMALNO VRIJEME SUŠENJA: 20 minuta

RJEŠAVANJE PROBLEMA

Proper Releaf® Restorative Position
Ensure the long end of the Restorative Leaf wraps around the retromolar area on the quadrant you are working on. To assist in evacuation, when a small amount of water and saliva build up behind the Restorative Leaf, have the patient slowly push their tongue against the Leaf while closing down on the Bite Block.
This will evacuate any remaining water that flows behind the Leaf.

Slaba veza
If you find the coupler becoming loose, it may be necessary to replace the O-ring.

Releaf® je registrovani zaštitni znak kompanije Innovative Dental Technologies, Inc.
Mora se prodavati isključivo od strane ili po narudžbi stomatologa.
Upotreba ovog uređaja je zabranjena u bilo koju drugu svrhu osim one koju preporučuje proizvođač.

Proizvođač

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (5)INNOVATIVE PRODUCT BRANDS™
Innovative Product Brands Inc. TM

IPB, Inc

ReLeaf-RESTORATIVE-Basic-Setup-Kit- (6)

FAQ

Are there any contraindications for using the ReLeaf Restorative system?

No known contraindications have been identified.

What precautions should be taken while using the product?

Risk of allergic reactions cannot be excluded for those predisposed to allergies.

How should loose connections be addressed?

If you find the coupler becoming loose, it may be necessary to replace the O-ring.

Dokumenti / Resursi

ReLeaf RESTORATIVE Osnovni komplet za pripremu [pdfUputstvo za upotrebu
RESTORATIVE Osnovni komplet za pripremu, RESTORATIVE, Osnovni komplet za pripremu, Komplet za pripremu, Komplet