U.S. Solid 5000g x 0.01g Analytical Balance

U.S. Solid Precision Lab Scale 5000g x 0.01g Analytical Balance Instruction Manual

Model: 5000g x 0.01g Analytical Balance with RS232 Interface

1. Uvod

This manual provides comprehensive instructions for the safe and efficient operation of your U.S. Solid Precision Lab Scale. Please read this manual thoroughly before use and retain it for future reference. This analytical balance is designed for precise weighing in laboratory, industrial, and educational settings, offering a capacity of 5000g with 0.01g readability.

2. Sigurnosne mjere

  • Always operate the scale on a stable, level, and vibration-free surface.
  • Izbjegavajte izlaganje vage ekstremnim temperaturama, vlazi ili direktnoj sunčevoj svjetlosti.
  • Ne preopterećujte vagu preko njenog maksimalnog kapaciteta od 5000 g.
  • Keep the weighing pan clean and free from debris.
  • Koristite samo isporučeni adapter za napajanje.
  • Izbjegavajte nagle udarce ili ispuštanje vage.

3. Postavljanje i instalacija

3.1 Raspakivanje

Carefully remove the scale and all accessories from the packaging. Verify that all components listed in the packing list are present. The package should include the scale, stainless steel weighing pan, and power adapter.

3.2 Postavljanje

Place the scale on a firm, level surface away from drafts, vibrations, and direct heat sources. Ensure adequate space around the scale for operation and ventilation.

3.3 Niveliranje skale

The scale must be perfectly level for accurate measurements. Use the adjustable feet at the bottom of the scale and the bubble level indicator to achieve proper leveling.

U.S. Solid Precision Lab Scale with bubble level indicator

Slika: Front view of the U.S. Solid Precision Lab Scale, showing the bubble level indicator on the left side of the display panel.

Diagram showing horizontal adjustment for leveling the scale

Image: Diagram illustrating how to adjust the leveling feet until the bubble is centered, ensuring proper horizontal adjustment for accuracy.

3.4 Power Connection and Warm-up

Connect the provided power adapter to the scale and then to a suitable power outlet. For optimal accuracy, allow the scale to warm up for at least 30 minutes after connecting to power or after a power loss before performing any measurements.

Video: This video demonstrates the initial setup of the U.S. Solid Precision Lab Scale, including connecting the power supply and the recommended 30-minute warm-up period for optimal performance.

4. Uputstvo za upotrebu

4.1 Osnovno vaganje

  1. Pritisnite ON/OFF dugme da uključite vagu.
  2. Wait for the display to show "0.00g".
  3. Place the item to be weighed on the center of the stainless steel pan.
  4. Očitajte stabilnu težinu prikazanu na LCD ekranu.

4.2 Konverzija jedinica

The scale supports 19 measurement units. Press the JEDINICA button to cycle through the available units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %).

Diagram showing 19 measurement units available on the scale

Image: Visual representation of the 19 different measurement units supported by the U.S. Solid Precision Lab Scale, including grams, kilograms, carats, and more.

4.3 Funkcija tare

Da biste oduzeli težinu posude, stavite praznu posudu na tavu i pritisnite dugme TARE/CAL button. The display will reset to "0.00g". You can then add the item to be weighed into the container, and only the net weight of the item will be displayed.

4.4 Piece Counting Function

The scale can count pieces of uniform weight. To use this function:

  1. Place asample quantity (e.g., 10, 20, 50, 100, 150, 200, 250, or 500 pieces) of the items to be counted on the pan.
  2. Pritisnite i držite COU dugme dok se na ekranu ne pojavi "PCS".
  3. Koristite JEDINICA dugme za odabir sample quantity that matches the number of items on the pan.
  4. Pritisnite TARE/CAL to confirm. The display will now show the number of pieces.
  5. Add more items to the pan, and the scale will display the total count.
Image showing piece counting function with various sample sizes

Image: Illustration of the piece counting feature, demonstrating how to set sample sizes (10, 20, 50, etc.) for accurate item counting.

4.5 Podešavanje pozadinskog osvetljenja

To adjust the backlight behavior:

  1. Pritisnite COU button to choose between C0-OFF, C0-1, or C0-5.
  2. C0-OFF: Backlight is always on.
  3. C0-1: Backlight will turn off after 1 second of inactivity.
  4. C0-5: Backlight will turn off after 5 seconds of inactivity.
  5. Konačno, kliknite TARE until the program ends and returns to 0.00.

Video: This segment demonstrates how to adjust the backlight settings of the scale, allowing users to choose between always-on, 1-second, or 5-second inactivity timeouts for the backlight.

4.6 Unit Package Locking

To lock specific unit packages for quick access:

  1. Pritisnite i držite COU button. Release it when "C0-OFF" is displayed.
  2. Kliknite TARE to choose C06-1.
  3. Pritisnite COU button to cycle through C06-1, C06-2, up to C06-16, each representing a different combination of units.
  4. Konačno, kliknite TARE until the program ends and returns to 0.00.

Video: This segment illustrates how to lock specific unit packages on the scale, allowing users to customize and quickly switch between preferred sets of measurement units.

5. RS232 Interface Communication

The U.S. Solid Precision Lab Scale features an RS232 interface for connecting to a computer or printer, enabling serial data exchange. This allows for data logging and integration with other systems.

Side view of the U.S. Solid Precision Lab Scale showing the RS232 port

Slika: Sa strane view of the U.S. Solid Precision Lab Scale, highlighting the RS232 interface port for external connectivity.

5.1 Key Output Mode Setting (C5-2)

In Key Output Mode, the scale sends weighing data to the connected device only when the JEDINICA dugme je pritisnuto.

  1. Pritisnite i držite COU button. Release it when "C0-OFF" is displayed.
  2. Kliknite TARE to choose C5-0.
  3. Pritisnite COU button to choose C5-2.
  4. Na kraju pritisnite TARE button to enter into key output mode.
  5. To output data, open "Uartassist" software (or similar serial terminal software), set Serial Port to COM3, Baud Rate to 9600, Data Bits to 8, Parity to None, and Stop Bits to One. Click "Open serial port".
  6. Svaki put kada pritisnete JEDINICA button on the scale, the computer will output the current weighing data.

Video: This segment demonstrates how to set the scale to "Key Output" mode (C5-2) for USB interface testing. It shows connecting the scale to a computer and manually triggering data output by pressing the UNIT button.

5.2 Continuous Output Mode Setting (C5-0)

In Continuous Output Mode, the scale continuously sends weighing data to the connected device without requiring manual input.

  1. Pritisnite i držite COU button. Release it when "C0-OFF" is displayed.
  2. Kliknite TARE to choose C5-0.
  3. Na kraju pritisnite TARE button to enter into continuous output mode.
  4. With "Uartassist" software (or similar serial terminal software) open and configured (COM3, 9600 baud, 8 data bits, no parity, one stop bit), click "Open serial port". The weighing results will be continuously output on the computer.

Video: This segment demonstrates how to set the scale to "Continuous Output" mode (C5-0) for USB interface testing. It shows the scale continuously sending weighing data to a connected computer.

6. Kalibracija

Regular calibration ensures the accuracy of your scale. The scale supports manual calibration.

  1. Ensure the scale is on a stable surface and has completed its warm-up period.
  2. With the scale displaying "0.00g", press and hold the TARE/CAL button until "CAL" appears.
  3. Na ekranu će se zatim prikazati trepćuća vrijednost težine (npr. "2000 g"). To je potrebna kalibracijska težina.
  4. Carefully place the specified calibration weight onto the center of the pan.
  5. Wait for the display to stabilize and show "PASS" or return to "0.00g".
  6. Remove the calibration weight. The scale is now calibrated.
Diagram showing calibration process for the scale

Image: Diagram illustrating the calibration process, emphasizing the need to recalibrate after moving the balance or before use, and to hold the TARE/CAL button for instructions.

7. Održavanje i njega

  • čišćenje: Obrišite vanjski dio vage mekom,amp cloth. Do not use abrasive cleaners or immerse the scale in water. The stainless steel pan can be removed for separate cleaning.
  • Skladištenje: When not in use for extended periods, store the scale in a dry, dust-free environment.
  • inspekcija: Povremeno provjeravajte kabel za napajanje da li ima oštećenja.

8. Otkrivanje problema

ProblemMogući uzrokRješenje
Vaga se ne uključujeNo power, faulty adapter, or power outlet issue.Check power connection, try a different outlet, ensure adapter is functional.
Netačna očitavanjaNot level, drafts, vibrations, not warmed up, needs calibration, or overloaded.Level the scale, move to a stable location, allow warm-up, perform calibration, ensure weight is within capacity.
Na ekranu se prikazuje "O-Ld"Preopterećenje.Remove excess weight immediately to prevent damage.
Greška u RS232 komunikacijiIncorrect port settings, cable issue, or wrong output mode.Verify COM port, baud rate, data bits, parity, and stop bits. Check cable connection. Ensure correct output mode (Key or Continuous) is selected.

9. Specifikacije

  • Brand: US Solid
  • Model: 5000g x 0.01g Analytical Balance
  • Kapacitet: 5000 grama (5 kg)
  • Čitljivost: 0.01 Grams (10 mg)
  • Vrsta prikaza: LCD sa pozadinskim osvetljenjem
  • Interfejs: RS232
  • Mjerne jedinice: 19 units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %)
  • Prijave: Check-weighing, Tare, Piece Counting, Manual Calibration, Overload Protection
  • Materijal: ABS kućište i posuda od nerđajućeg čelika
  • Dimenzije proizvoda: 8.9 x 8.7 x 3.2 inča
  • Težina artikla: 2.9 funti
  • Snaga: AC adapter (uključen)

10. Garancija i podrška

For warranty information and technical support, please refer to the warranty card included with your product or visit the official U.S. Solid webSačuvajte račun kao dokaz o kupovini.

Kontakt informacije: Molimo pogledajte službene informacije proizvođača website or contact details provided in the product packaging for support.

Povezani dokumenti - 5000g x 0.01g Analytical Balance

Preview Uputstvo za upotrebu analitičke vage US Solid USS-DBS
Comprehensive operating manual for the U.S. Solid USS-DBS series analytical balance, covering features, specifications, installation, usage, maintenance, and troubleshooting.
Preview U.S. Solid USS-DBS82 Precision Balance Instruction Manual
Comprehensive instruction manual for the U.S. Solid USS-DBS82 Series Precision Balance, covering safety precautions, design, operation, technical specifications, and contact information.
Preview Uputstvo za upotrebu analitičke vage US Solid USS-DBS
Sveobuhvatni priručnik za upotrebu analitičke vage US Solid USS-DBS, s detaljnim specifikacijama, instalacijom, radom, unutarnjim postavkama, održavanjem, rješavanjem problema i zakonskim ograničenjima upotrebe.
Preview Uputstvo za upotrebu analitičke vage US Solid USS-DBS
Comprehensive operating manual for the U.S. Solid USS-DBS series analytical balances, covering features, specifications, installation, usage, calibration, maintenance, troubleshooting, and RS232 interface.
Preview U.S. SOLID Analytical Balance Manual: Precision Weighing, Density & Dynamic Functions
Comprehensive instruction manual for U.S. SOLID 0.1 mg Analytical Balances, covering models like USS-DBS62 and USS-DBS63. Learn about precision weighing, density determination, dynamic weighing, calibration, operation, and troubleshooting for laboratory and industrial applications.
Preview U.S. Solid USS-DBS Digital Balance Operating Manual
Comprehensive operating manual for U.S. Solid USS-DBS series digital balances, covering specifications, installation, usage, maintenance, and troubleshooting. Learn how to calibrate, tare, and use piece counting functions.