U.S. Solid USS-DBS64-500

U.S. Solid 500g x 0.001g Lab Analytical Balance Instruction Manual

Model: USS-DBS64-500

1. Uvod

This manual provides essential instructions for the proper setup, operation, maintenance, and troubleshooting of your U.S. Solid 500g x 0.001g Lab Analytical Balance. Designed for high-precision weighing, this balance offers reliable results with 0.001g (1mg) accuracy and a maximum capacity of 500g. It features multiple unit conversions, parts counting, percentage weighing, and manual calibration capabilities.

Please read this manual thoroughly before using the balance to ensure accurate measurements and safe operation.

2. Sigurnosne mjere

  • Always operate the balance on a stable, level, and vibration-free surface.
  • Avoid placing the balance near strong electromagnetic fields, direct sunlight, or rapid temperature changes.
  • Do not overload the balance beyond its maximum capacity of 500g.
  • Keep the balance and its components clean and free from dust and corrosive substances.
  • Use only the provided power adapter or a compatible 9V battery.
  • Do not attempt to disassemble or repair the balance yourself. Contact qualified service personnel if needed.

3. Šta se nalazi u kutiji

Carefully unpack all components and check against the following list:

  • Scale Body x1
  • Scale Pan x1
  • Kabl za napajanje x1
  • Calibration Weight x1
  • Uputstvo za upotrebu x1

4. Proizvod je gotovview

Familiarize yourself with the main components of your analytical balance.

Front view of the U.S. Solid Lab Analytical Balance with windshield and display showing 0.000g.
Slika 4.1: Front view of the balance. Shows the main display, control buttons, weighing pan, and windshield.
Side view of the U.S. Solid Lab Analytical Balance, highlighting the compact design.
Slika 4.2: Side view of the balance, illustrating its compact form factor.
Pozadi view of the U.S. Solid Lab Analytical Balance, showing the power input.
Slika 4.3: Pozadi view of the balance, including the power input port.
Close-up of the bubble level on the U.S. Solid Lab Analytical Balance for horizontal adjustment.
Slika 4.4: Detail of the integrated bubble level for precise horizontal adjustment.
Eksplodirao view of the U.S. Solid Lab Analytical Balance components with dimensions.
Slika 4.5: Eksplodirao view showing all components and their dimensions.

5. Podešavanje

5.1 Raspakivanje i početni pregled

Carefully remove the balance from its packaging. Inspect all components for any signs of damage. If any damage is found, do not use the balance and contact customer support immediately.

5.2 Postavljanje

Place the balance on a firm, level, and stable surface. Ensure the location is free from vibrations, drafts, direct sunlight, and rapid temperature fluctuations, which can affect measurement accuracy.

5.3 Izravnavanje ravnoteže

The balance must be perfectly level for accurate readings. Use the adjustable feet at the bottom of the balance to center the bubble in the integrated level indicator (refer to Figure 4.4).

Image showing the horizontal adjustment process with the bubble level.
Slika 5.1: Horizontal Adjustment. Adjust leveling feet until the bubble is centered. Proper leveling is essential for accuracy.

5.4 Priključak za napajanje

Connect the provided power cord to the balance and a suitable power outlet (100-240V). Alternatively, the balance can be powered by a 9V battery for portable use.

5.5 Warm-up Time

After connecting power, allow the balance to warm up for at least 30 minutes before taking measurements. This stabilizes the internal components and ensures optimal accuracy.

6. Uputstvo za upotrebu

6.1 Uključivanje/isključivanje napajanja

  • Pritisnite ON/OFF button to turn the balance on.
  • Pritisnite i držite ON/OFF button to turn the balance off.

6.2 Osnovno vaganje

To perform a basic weight measurement:

  1. Ensure the balance is level and has completed its warm-up.
  2. Pritisnite TARE/NULA button to set the display to 0.000g.
  3. Pažljivo postavite predmet koji treba vagati na sredinu vage.
  4. Wait for the reading to stabilize. The stable weight will be displayed.

Klip 6.1: Demonstration of basic weighing, unit conversion, and percentage function on the U.S. Solid Lab Analytical Balance.

6.3 Funkcija tare

Funkcija tare vam omogućava da oduzmete težinu posude, tako da se mjeri samo neto težina sadržaja.

  1. Stavite praznu posudu na posudu za vaganje.
  2. Pritisnite TARE/NULA dugme. Prikaz će se resetovati na 0.000 g.
  3. Dodajte materijal koji treba izvagati u posudu. Na ekranu će se prikazati neto težina materijala.

6.4 Konverzija jedinica

The balance supports multiple weighing units. To change the unit:

  • Pritisnite JEDINICA button repeatedly to cycle through available units: g (grams), ct (carats), oz (ounces), ozt (troy ounces), lb (pounds), gn (grains), t (metric tons), dwt (pennyweights), kg (kilograms), dr (drams), tlT (Taiwan Tael), N (Newtons), and % (percentagi).
  • Odabrana jedinica će biti prikazana na ekranu.
Image showing the various measuring units available on the balance.
Slika 6.1: Available Measuring Units and Piece Counting options.

6.5 Parts Counting (PCS)

Ova funkcija vam omogućava da prebrojite veliki broj identičnih predmeta po težini.

  1. Place a known quantity of items (e.g., 10, 20, 50, 100, 150, 200, 250, 500) on the pan.
  2. Pritisnite PCS dugme.
  3. Select the corresponding sample quantity using the JEDINICA dugme.
  4. Pritisnite PCS again to confirm. The balance is now in counting mode.
  5. Add more items, and the display will show the total count.

6.6 Procenattage Vaganje

Procenattage function is useful for formulation and quality control, allowing you to measure a sample's weight as a percentage referentne težine.

  1. Stavite referencu sample na posudi za vaganje.
  2. Pritisnite JEDINICA button until '%' is displayed. The reference weight will be set as 100%.
  3. Uklonite referencu sample.
  4. Postavite sljedeće samples on the pan. Their weight will be displayed as a percentage referentne težine.

6.7 Kalibracija

Regular calibration ensures the accuracy of your balance. It is recommended to calibrate the balance upon initial setup, after relocation, or if you suspect inaccurate readings.

  1. Ensure the weighing pan is empty and the balance is stable.
  2. Pritisnite i držite CAL dugme dok se na ekranu ne pojavi 'CAL'.
  3. The display will then show the required calibration weight (e.g., 500.000g).
  4. Carefully place the specified calibration weight (provided) onto the center of the weighing pan.
  5. Wait for the balance to automatically calibrate. Once complete, the display will show 'PASS' or return to normal weighing mode.
  6. Uklonite kalibracijski uteg.
Image showing the control panel buttons including ON/OFF, UNIT, PCS, CAL, and TARE/ZERO.
Slika 6.2: Control Panel Buttons. Key functions include ON/OFF, UNIT, PCS (Parts Counting), CAL (Calibration), and TARE/ZERO.

7. Održavanje

7.1 Čišćenje

To maintain accuracy and prolong the life of your balance:

  • Uvijek isključite napajanje prije čišćenja.
  • Obrišite vanjski dio vage mekom,amp tkanina. Po potrebi koristite blagi deterdžent.
  • Ne koristite abrazivna sredstva za čišćenje, rastvarače ili uranjajte vagu u vodu.
  • Clean the stainless steel weighing pan and the windshield regularly to prevent accumulation of dust or residues.

7.2 Skladištenje

When not in use, store the balance in a clean, dry environment, away from extreme temperatures and humidity. Keep the windshield in place to protect the weighing mechanism.

8. Otkrivanje problema

This section addresses common issues you might encounter with your analytical balance.

ProblemMogući uzrokRješenje
Netačna ili fluktuirajuća očitanja
  • Balance not level
  • Vibrations or drafts
  • Promjene temperature
  • Potrebna je kalibracija
  • Preopterećenje
  • Adjust leveling feet (Figure 5.1)
  • Relocate balance to a stable environment
  • Allow balance to acclimate to room temperature
  • Perform calibration (Section 6.7)
  • Ensure weight is within 500g capacity
Display shows 'Err' or other error codesInternal fault or overload condition.
  • Remove all items from the pan.
  • Turn off the balance, wait 30 seconds, then turn it back on.
  • Ako greška i dalje postoji, kontaktirajte korisničku podršku.
Balance does not power on
  • Nema napajanja
  • Faulty power adapter/battery
  • Provjerite priključak kabla za napajanje i utičnicu.
  • Replace 9V battery if using battery power.
  • Ako je moguće, testirajte adapter za napajanje s drugim uređajem.
Cannot detect small weight changes (e.g., 10-15mg)
  • Environmental interference (drafts, vibrations)
  • Balance not properly calibrated
  • Warm-up time insufficient
  • Ensure stable environment, use windshield.
  • Perform calibration (Section 6.7).
  • Allow sufficient warm-up time (Section 5.5).

9. Specifikacije

Technical specifications for the U.S. Solid 500g x 0.001g Lab Analytical Balance.

FeatureDetalj
BrandUS Solid
Broj modelaUSS-DBS64-500
Ograničenje težine500 grama
Preciznost očitavanja0.001 Grams (1mg)
Display TypeLCD (pozadinsko osvetljenje)
Vrsta vageBalans Scale
mjerne jediniceg, ct, oz, ozt, lb, gn, t, dwt, kg, dr, tlT, N, %
NapajanjeAC (100-240V) / DC (9V Battery)
Vrsta materijalaAkrilonitril butadien stiren (ABS), nehrđajući čelik
Težina artikla3.64 funti
Uključene komponenteScale Body, Scale Pan, Power Cord, Calibration Weight, Manual

10. Garancija i podrška

U.S. Solid products are manufactured to high-quality standards. For information regarding product warranty, please refer to the warranty card included with your purchase or visit the official U.S. Solid website.

For technical support, troubleshooting assistance, or to inquire about replacement parts, please contact U.S. Solid customer service. Contact details can typically be found on the product packaging or the official brand website.